Filling the CCD dewar
The CCD needs to be cooled to a temperature of -130 degrees C. The temperature
can be checked from within voodoo. The dewar has
a hold time of about 24 hours, and is cooled with liquid nitrogen.
The LN2 is stored in the brown 35-liter
storage dewars in the dome (Dawn Clemons can be contacted at
dclemons@noao.edu to ask her to fill one of these
before the observing run).
Because the geometry of the dewar makes leakage of liquid N2 possible for certain telescope
orientations, we recommend that the dewar be filled at the end of the
night or at lunchtime rather than just before observing starts to
minimize wastage of liquid N2.
Filling instructions:
- Wheel the 35 liter storage dewar which is currently in use under
the CCD dewar on the telescope.
- The long thin rod (stinger) stored on the side of the storage
dewar should be already connected to its liquid output valve. Take the
stinger assembly and carefully insert the stinger into the fitting at
the bottom of the CCD dewar. There is an obstruction part way up the
pipe and the stinger tip needs to be well centered in the pipe to get
past it. Screw the stinger fitting closed; hand tight is sufficient.
- Note that especially talented astronomers have managed to place
the dewar so that the flexible filling tube, which freezes during
filling, is underneath the liquid N2 valve and so prevents closing of the
valve. Make sure that it will be possible to close the valve once the
hose freezes solid!
- Especially when the storage dewar is full, the gas pressure
inside it may not be sufficient to push the liquid N2 out. About 10
psi (read on the gauge on the top of the dewar) is optimum. The dewar
safety valves will let go at 12 psi. If the storage dewar pressure is
less than about 5 psi, you will need to add dry N2 to increase the
pressure. There are two dry N2 outlets on the S pier of the telescope.
Use the tubing connected to the outlet without a flow-meter attached.
Connect this tubing to the vent valve on the storage dewar. Turn on
the dry N2 and and open the storage dewar valve. (Dry N2 pressure on
the tank downstairs should be about 15 psi).
- Open the liquid valve on the storage dewar (move handle
vertical) so that liquid N2 flows into the CCD dewar. Fill until
liquid N2 rather than gas flows from the overflow spout on the stinger
connection. This should take just a few minutes or so.
- Turn off the liquid N2 valve on the storage dewar, the vent
valve on the storage dewar, and the dry N2 on the S pier and the N2
cylinder downstairs (if storage
dewar was pressurized).
- Wait until the fill assembly warms, and then unscrew connection
to CCD dewar, remove stinger and replace in tube on the side of the
storage dewar. If you are in a hurry, you can use the hair dryer
gently to warm the fill assembly before disconnecting it.
Swapping the fill heads on the LN2 tanks
If one tank runs out, you need to swap the fill head
to the other tank.
(If both tanks have run out, you didn't plan ahead!)
Before you do so, you need to make sure the empty tank
has been depressurized. If it is absolutely empty, you
can just let it vent via the fill tube, but if it's not
empty, then this will spray LN2 everywhere (amusing, perhaps,
but not recommended...)
There are 3 ways to relive the pressure, none of them ideal.
We need to add a bleed valve on the gas line at the head of the tank
use a snap fitting like we use for the compressor of the N2 gas connection.
-
The quickest way to relieve pressure is to disconnect the N2 gas line
and open the gas valve (slowly). However this requires a wrench to
undo the fitting, and it will leak if we routinely undo it.
-
Another way is to hold open one of the spring loaded safety valves
there are 2 approx 3/8" diameter silver fittings with brass end plug.
You often hear gas coming out of these when you have the N2 gas
line connected. (if there is much pressure you finger will get
cold before it bleeds off.
-
The default way if you didnt think about it.
MAKE SURE THE WIRE CABLE IS CONNECTED TO THE RING
with the dog lead like hook.
Slowly unscrew the black knob, on the clamp that holds the head
to the storage dewar. Gas should bleed out around the Oring as
the clamp is loosened. I think there is approx 1" of thread
for the knob to unscrew, the seal should have broken before
the knob unscrews halfway.
Assuming the pressure gauge reads zero, undoo the black knob
all the way off the thread.
Open the clamp by hand, so that it can be fiddled/removed.
You can now just lift up the head out of the dewar.
Be careful of the long pipe that extends to the bottom of the dewar,
if its cold its somewhat brittle.
Reverse the procedure to install.
Table of contents
5. Computers at the Burrell Schmidt
7. CCD Acquisition